Instruction User Manual Of Handheld 3 Liter Portable Battery Powered Oxygen Concentrator

Handheld Portable Battery Powered Oxygen Concentrator

Handheld 3 Liter Portable Battery Powered Oxygen Concentrator Making Oxygen Concentrador De Oxigeno
 

Instruction Manual of Handheld Portable Battery Powered Oxygen Concentrator

Model: 3 Liter Portable Battery Powered Oxygen Concentrator
Please read the instruction manual carefully before using the product.
 
WebSite: https://dedakjoxygenconcentrators.com/
Mail Address:  dedakj@yahoo.com

 

 

Table of Contents

Chapter1: Introduction
Chapter2: Description of theXM-868N Portable Concentrador De Oxigeno
Chapter 3:  Operating Instructions
Chapter 4:  Cautions and Warnings
Chapter 5:  Specifications
Chapter 6:  Accessories
Chapter 7:  Care and Maintenance
Chapter 8:  Warranty Statement

 

Introduction

In order to reduce the risk of personal injury,ensure that the following precautions are not disregarded.
1.The product is used on a health care basis by users requiring supplemental oxygen,instead of medical purpose. 
2.If you begin to feel ill or are experiencing discomfort while using this device, stop using and consult the physician immediately.
3.Ensure this user manual is being read carefully and completely understood by users before use.
4.The expected life for the device is 3years, applications and maintenance may affect its intended life.
5.The XM868N is not designed or specified to be used in conjunction with a humidifier, nebulizer or connected to any other equipment.
6.The device produces enriched oxygen gas which accelerates combustion,do  not allow smoking or open flames near the product while in use.
7.This device is NOT INTENDED to be life sustaining or life supporting.
8.Do not expose to water precipitation or snow.
9.Never leave the device in an environment which can reach high temperatures and high humidity 
such as an unoccupied car in high temperature environments. This could damage the device.
10.Donot disassemble the product or any of the accessories as any modification of them 
may damage the device or impair equipment performance, and will void your warranty.
11.Do not submerse the XM868N in water or any other conductive liquid. If  they occur, disconnect the power supply immediately.
12.Make sure to unplug the power supply while not using it.
13.Do not obstruct air intake or exhaust when operating the device.
Blockage of  air circulation or proximity to a heat source may lead to internal heat buildup and impair equipment performance or may even spark a fire.
14.After continuous use for 120 minutes, stay in standby mode for 20-30  minutes for heat dissipation.
15.Do not place other devices near,above or under the concentrator. 
For necessary occasions,ensure it is operated under using configurations after observations and approvals.

 

Description of the XM-868N Portable Concentrador De Oxigeno

 Specifications

Items

Parameters

Product

Portable Oxygen Concentrator (With Rechargable Battery and backpack)

Model

XM-868N Portable Concentrador De Oxigeno

Method

Vacuum Pressure Swing Adsorption(VPSA)

Dimensions

208mm*78mm*206mm

Weight

Approximate 2KG(Concentrator only)

Noise

≤50db(A)Test 1 meter away from the device

 

Flow/Oxygen Concentration

90% concentration in l setting,equivalent to lL  

70% concentration in 2 setting,equivalent to 2L  

50% concentration in 3 setting,equivalent to 2L

Maximum Outlet Pressure

≥83KPa

Negative Oxygen lons ≥300K/m³(Electron  Emission)
Power Supply AC100-240V50/60Hz
Power 20W
Battery Duration 2H
Battery Charging Time 2H
  Environment Ranges Intended for Use Temperature:5℃-40℃ Humidity:≤10%-90%,non-condensing Atmospheric Pressure Range:54hpa~1060hpa

 

Accessories

Packing List
Number Item Quantity
1 XM-868N 1
2 Carry Bag 1
3 Nasal Cannula 1
4 Power Adapter 1
5 Car Cigarette Lighter 1
6 User Manual 1
7 Remote Control 1

Operating Instructions

1.Power  Supply
Select the suitable power connections subject to the applications: Press and hold the power button for 2seconds (Image 2), when the built in battery is being used only. A single short beep will sound after the LED is on.Power display will indicate the power condition and the device is on.
  When using the car power adapter, ensure that the connections between the power adapter and power cord are firm enough(Image 1),
if the power indicator of the equipment shows it is being charged, indicating  the  power connection is all  set. 

When using the AC power adapter,ensure that the connections between the power adapter and power cord are firm enough(Image 1),if the power indicator of the equipment shows it is being charged,indicating the power connection is all set.

 

2.0xygen  Generation
After the concentrator is power on, press the oxygen generation
Button to enter the 1st setting(Image 3). A warm up period of up to 2
minutes will initiate till the settings are completed.Use it with the nasal
cannula  accordingly.
Note: During the warm-up period, the equipment will automatically
exhale and empty the internal oxygen, hence the oxygen concentration
may not reach the standard at this time.

 

3.Nasal  Cannula

Connect the nasal cannula tubing to the nozzle fitting (image  4).
Make sure the connection is firm without leakage.Do not kink and
obstruct the nasal cannula, as it may cause alarms of the device and
affect the oxygen delivery.
Position the nasal cannula on your face(Image 5) properly and
inhale.

 

4.Flow  Setting

Under the oxygen generation mode, press the oxygen generation
button to select the settings from 1 to 3(Image 3).

 

5.Power off

When the device is working, press and hold the power button for 2  seconds, a  beep will sound and it indicates power off.
If the device is connected to an adapter, car charging cable etc.during standby mode, the power indicator on the display will show it is  being  charged.
When the battery  is fully charged, disconnect the adapter to power off manually.
If the  device  is  powered  off  when  using  battery  only,  the  device  will  be  completely powered off and the standby interface will not appear. 
Under the standby mode,it will automatically power off after 1 minute if there is  no operation.

 

Effective way to use the concentrator

The product is designed for continuous use. For better effects, it should be used 30 minutes per day regularly. Schedule for use is optional.

 

Cautions and Warnings

Warnings

Incorrect operations may lead to death or heavy injury

 

Do not use when the adapter plug is damaged or power socket is loose. Danger of fire, short-circuit and electric shock.

 

Keep cords away from children if it is for their use. To avoid danger of choking or strangulation hazard.

 

Do not smoke while in use.

Smoking nearby is not allowed when the device is on.

 

Do not use nebulizer or spray any volatile gases near the concentrator. It may cause fire if they combine with oxygen.

 

Do not place the device close to gasoline supply. It may lead to failure of performance or fire.

 

Do not repair, disassemble or modify the equipment if you are not technical personnel. Failure of performance of the device could occur.

Do not apply multiple adapter plugs to one power socket. Doing so may create a safety hazard.

 

Do not drag the plugs or contact them with wet hands. Do not drag the plugs or contact them with wet hands.

 

Do not damage the cord, such as kink, modify or place heavy objects over it. The cord may be damaged,fire and electric shock may occur.

 

Do not use this device for other purposes than oxygen supplement. This could cause electric shock.

 

 

Cautions

Incorrect operations may lead to injury or property loss.

Do not block the intake vent which is below the device. Oxygen concentration will be low or the oxygen flow is not sufficient if it occurs.

Avoid direct sunlight or humid places.

To prevent the product from damage and electric shock or fire.

Unplug the power supply from sockets while not using.

Deterioration of the plug may lead to electric shock or fire.

Ensure that the nasal cannula is being used appropriately. Do not damage the nasal cannula,no stepping on the cannula with feet or furniture.Do not trip by it as well

Do not allow open flames near the product while in use. The device produces enriched oxygen gas which accelerates combustion.Lghters or matches are prohibited to use nearby.

Do not apply the concentrator in inclined places. It may cause falling off the device which leads to damage and impairment.

Do not use the device in the bathroom or sauna. Doing so may cause water damage,and dampness may lead to electricleakage and damage.

After continuous use for 120 minutes, stay in standby mode for 20-30 minutes.

For heat dissipation.

 

Accessories

Packing List

Number

Item

Quantity

1

XM-868N Portable Concentrador De Oxigeno

1

2

Carry Bag

1

3

Nasal Cannula

1

4

Power Adapter

1

5

Car Cigarette Lighter

1

6

User Manual

1

7

Remote Control

1

 

 

Care and Maintenance

1.Filter Cleaning and Replacement

(1) Turn off the device, remove the screws, and remove the filter case.
(2)Take the filter out (Note: The filter will appear black after long-term use).
(3) Rinse the filter in clean water, dry it before reuse.
(4)Install the filter in correct direction,and install the filter case,fix it with screws.
Note:
(1)Maintenance of the filter shall be done every 3-6 months.
(2) Do not use detergents,alcohol and other high-concentration liquids to clean, as it will cause cracks or discoloration of the filter.
(3) Do not use the hot air of a hair dryer to dry the filter, which may cause deformation of the filter.
 

2.Sieve Replacement and Maintenance

(1)When the molecular sieve needs to be replaced, contact the appointed seller in time.(When the device is used for 3000 hours in total or more than 2years,replace the sieve in time).
(2)Long-term storage may shorten the life of the sieve;
(3) Longterm operation in a humid environment may shorten the service life of molecular sieve.
Note: Do not disassemble the equipment to replace the molecular sieve privately,
only the equipment supplier or qualified maintenance engineer is qualified to disassemble and replace it.

 

3.Cannula  Replacemen

Users can purchase the nasal cannula according to your needs, but ensure that the following adjustments are fully satisfied:
(1) Contact your equipment provider and purchase under the guidance of the professionals.
(2) Purchase from regular distributors.
(3)The rated flow rate of oxygen through the nasal cannula is5 Liters per minute, and the total length is between land 7.5meters.
(4)Make sure it can fit properly with the oxygen nozzle of the device without air leakage.
Position the nasal cannula correctly on your face and use it according to instructions. 

 

4、Battery Care

Battery loss:500 cycles charge and discharge capacity will drop to 80%

Regular charging and maintenance(Approximately I month) when not in use fora longtime,it is recommended to keep the power at 3-4 LEDs Light when storing.

5.Case Cleaning

Make sure the concentrator is powered off and unplug the power supply from the socket before cleaning.
Keep the device away from dampness or water damage,
once it occurs,it will cause malfunction or shutdown of the equipment,
and increase the risk of electric shock.
Do not pour the liquid directly on the device,chemical detergent used on the plastic case of the device may damage it.  
Detergents contain but not limited to the following lists: Alcohol and alcohol products,high concentration chlorinated solutions,oil products(Clorox),etc.
Be cautious that liquid is not permitted to apply in the internal equipment,
and pay special attention to ensure that there is no dust, water,or other particles at the oxygen nozzle.
Do not clean it with solvents or other flammable  and explosive volatile substances . 

 

6.Maintenance Troubleshooting

Number Problem Possible    Caus Recommended       Solution
        1         Concentrator does not power on   When On/Of button is pressed Press and hold the power button for 2 seconds to light up the screen,short press the oxygen generation button to start up.
Battery is dead Plug the adapter and charge
AC Power supply is not connected properly Check power supply connection
  Malfunction Contact your equipment provider  if the above operations are not helpful.
    2     No oxygen Concentrator is not powered on Power the  device
Cannula is kinked or obstructed Check cannula and its connection to concentrator nozzle
Malfunction Contact your equipment provider
    3     Device over heat High    indoor temperature

Ensure that the connection 

between the external air and internal exhaust is smooth

Obstruction of the exhaust vent at the bottom
Long-time  use Power off the device after it is being used forabout2hours
  4 Abnormal smell from exhaust vent or device Use time of the nasal cannula is overly long

Clean the cannula regularly and replacement is 

recommended when it is used for too long

 

 

DEDAKJ Oxygen Concentrator

Official Site:

https://dedakjoxygenconcentrators.com/

 

 dedakj portable oxygen concentrator medical oxygen machine

 