MIni Portable Oxygen Concentrator
Instruction Manual of KPS Series Portable oxygen concentrator
Contents
Product Diagram
Product Specifications
Interface Introduction
Directions For Use
Maintenance And Cleaning
Warranty Description
Common Problems And Solutions
Packing List
Alarm Information
Repair Information
PRODUCT DIAGRAM
Operation main screen
PRODUCT SPECIFICATION
Product Name: portable oxygen concentrator
Product size:L*W*H :15.0cm*7.5cm*16.0cm
Product weight:Machine weight : 1.1Kg , Battery weight : 0.3Kg
Interface:2.8"monitor
Operating noise:≤60dB(A )( position 5 )
Audio alarm:≥45dB(A)
Sea level: 5500M normal use
Warm-up time:3min
Oxygen concentration:≥90% ( At all gears and respiratory rates )
Gear setting: 1, 2, 3, 4,5
Respiratory rate:10-40 BPM
Respiratory sensitivity:≤0.12 cmH20
Max outlet pressure:100KPa+20%
AC Power Source:AC100-240V 50/60 Hz 6OVA MAX
Internal battery:14.8VDC 4A Max
Battery duration:1.5~4 hours
Battery charging time:2.5h
Suitable environment: (1) Temperature : 5~ 40°C; (2) Humidity : 10%~90% , non condensing state; (3) Atmospheric pressure :540hPa~1060hPa
*The above data are tested by the laboratory, and the final interpretation right belongs to the company
INTERFACE INTRODUCTION
Settings interface
Long press ON/ OFF
Click on the sensitivity button to select the appropriate sensitivity.
Click on on the timing button to select timing.
Click the plus or minus button to select the appropriate flow.
DIRECTIONS FOR USE
1、Power supply
Select the appropriate power connection conditions based on the usage environment.
When only using batteries, install the dedicated battery for this device to the device battery interface.
When using the on-board power adapter, when the power indicator on the adapter lights up.
it indicates that the power connection is normal.
Required input terminal: DC 19v 65VA MAX.
When using the AC power adapter, connect the AC power adapter and the power cord firmly. When the power indicator on the adapter lights up, it indicates that the power connection is normal.
Required input terminal: AC 100-240V, current 2.0-1.0A, and frequency 50/60Hz.
Warning: Do not use lithium batteries that are not included with the product.
2、Startup preheating
Press the device ON key, and after hearing a "tick" boot prompt tone, enter the boot screen.
After the system starts, the device enters a warm-up state. The preheating time of the equipment is about 2 minutes. After the preheating is completed, when the equipment status meets the requirements for normal use, it can be worn and equipped with a nasal oxygen tube for normal use.
Note: During the warm-up period of the machine, the device will automatically inject oxygen to empty the internal air, At this time, the output flow and concentration of the device may not meet the standard, so do not use it.
Note: Do not connect the nasal oxygen tube during the warm-up period of the machine
3、Connecting the nasal oxygen tube
Connect the factory equipped nasal oxygen tube bell mouth to the metal air outlet of the equipment to ensure reliable connection and no air leakage. Note that the nasal oxygen tube should not be kinked or blocked to avoid causing equipment related alarms and affecting normal use.
Note: This device is a pulse oxygen supply method. To use this device, it must be connected to a nasal oxygen tube.
4、Adjusting the flow gear
To adjust the flow gear, please press the setting key to set the flow gear according to your needs
5、Turn off
When the device is working normally, press and hold the ON/ OFF key for about 3 seconds, and the device emits a "beep" prompt and enters the shutdown interface. The device automatically shuts down.
6、Other information
When the adapter is connected/ unplugged or the battery is connected/ unplugged, the device will emit a "tick" prompt, and the main interface of the device will also have relevant information to display the current power supply status.
When an abnormal state occurs, the abnormal information indicator light will light up, and the device will display abnormal information and issue an audible prompt. if you have learned the prompt information and no longer need a prompt tone, you can press the prompt tone switch to turn off the information prompt tone.
Note: When the alert tone is turned off, please pay close attention to the device status to prevent you from not being able to fully grasp the device operation without the alert tone
Note: It is prohibited to remove the battery during equipment operation.
MAINTENANCE AND CLEANING
1. Replacement of intake filter cotton
- 1.Shut down the device, disconnect the power supply and remove the battery
- 2.Remove the filter cotton bottom shell.
- 3.Take out the intake filter cotton. Note that the intake air filter cotton after long-term use mayabsorb and accumulate a large amount of dust. Please operate with care and take properprecautions to prevent inhalation of dust.
- 4.Install new filter cotton in the correct direction
- 5. Install the filter cotton bottom shell back.
- 6.Please properly dispose of the waste filter cotton
Note: Failure to install or replace the filter cotton for a long time may affect equipment performance or damage the equipment. Please replace the filter cotton in a timely manner, and the replacement cycle depends on the use environment and usage. Generally, it is replaced once a month.
2. Molecular sieve replacement and maintenance
- When the device prompts the need to replace the molecular sieve, please contact the designated seller in a timely manner.
- Long term storage may shorten the service life of molecular sieves.
- Long term operation in a humid environment may shorten the service life of molecular sieve.
Warning: Do not disassemble and replace molecular sieves without permission. Only equipment suppliers or qualified maintenance engineers can disassemble and replace molecular sieves.
3. Nasal oxygen tube replacement
Users can purchase nasal oxygen tubes as needed, but they need to ensure that the following conditions are fully met:
- Please contact the equipment provider or purchase under the guidance of a professional (Official Sore:https://dedakjoxygenconcentrators.com/ ).
- Please purchase from regular channels.
- The rated flow rate of oxygen in the nasal oxygen tube is 5 liters per minute, and the total length is between 1 and 7.5 meters
- It can fit well with the air outlet of this product without air leakage.
Please wear a nasal oxygen tube correctly and use it according to the instructions for using the nasal oxygen tube.
4. Battery replacement and maintenance
This device requires the use of a dedicated battery provided at the factory, with a standard battery provided at the factory. Users can contact a designated seller to purchase batteries as needed.
Battery life: 500 cycles of charge and discharge capacity reduced to 80%.
When not in use for a long time, please remove the battery from the device and pay attention to protecting the battery electrodes from contacting conductors such as metals to prevent fire and other hazards.
Please store the battery away from children to avoid danger.
Batteries that have not been used for a long time need to be regularly charged and maintained.
It is recommended that the battery power be kept at about 30% to 60% when stored.
Warning: Do not use lithium batteries that are not included with the product.
5. Service life
Oxygen generator host: 3 years
molecular sieve: 1 year
Battery: 500 full charges and discharges
Note: The service life in the table is the recommended value, and the actual service life will vary depending on the actual usage environment and actual usage, Please pay close attention to the equipment usage status and related alarm prompts.
6. Maintenance cycle
Air intake filter cotton: Once a month- User replacement (the original filter cotton can be purchased separately.)
Molecular sieve assembly: Once a year- Paid replacement (Replace as prompted by the device)
Battery: Once three months-Charge to 30% (No special requirements for normal use)
Note: The maintenance cycle in the table is the recommended value, and the actual maintenance cycle will vary depending on the actual usage environment and actual usage, Please pay close attention to the equipment usage and related alarm prompts.
7. Equipment cleaning
Before cleaning, please ensure that the machine is turned of, unplug the power supply from the socket, and remove the battery.
The power off sequence should be to first disconnect the power plug on the wall and then disconnect the plug on the machine side.
Do not let the equipment get wet or enter water, Once this happens, it will lead to equipmen failure or shutdown, and it will also increase the risk of electric shock.
Do not pour liquid directly onto the machine. Chemical cleaning agents used on the plastic exterior of the machine may damage the plastic housing of the equipment. These cleaning agents include, but are not limited to, the following lists: alcohol and alcohol products, high concentration chlorinated solutions (chloroethane), and oil products (Kolol).
Be careful not to allow liquid to enter the interior of the machine. Pay special attention to ensuring that the oxygen outlet is free of dust, water, or other particles.
Do not use organic solvents or other flammable and explosive volatile substances for cleaning.
Please ensure that the equipment has been completely dried before use.
Warning: Before disposal, remove batteries, etc., and contact relevant departments to properly dispose of equipment and accessories in accordance with local government regulations.
WARRANTY DESCRIPTION
- The oxygen generator host (excluding consumables) is warranted for one year from the date of purchase.
- All maintenance services must be carried out by a designated maintenance service center.
- During the warranty period, faults not caused by human factors are warranted free of charge.
- After the warranty period, maintenance costs are charged.
- If damage is caused due to improper use, paid maintenance services shall be provided.
- Consumable nasal oxygen tubes and filter cotton are not covered by the warranty.
COMMON PROBLEMS AND SOLUTIONS
Problem |
Possible causes |
Recommended Solution |
Device can not be opened |
Battery not installed correctly |
Remove the battery and reinstall it |
|
Dead battery |
Insert the battery and connect the adapter to charge the battery |
|
Poor AC power connection |
Check the power connection |
|
Equipment failure |
If the above operation cannot solve the problem, please contact your equipment supplier |
No oxygen output |
Device not turned on |
Turn on the oxygen generator |
|
Twisted or blocked nasal oxygen tube |
Check the nasal oxygen tube and its connection to the oxygen outlet |
|
Equipment failure |
Contact the equipment supplier |
The following states are not faulty
The exhaust port is a heat dissipation port, When operating for a long time or the ambient temperature is high, the temperature of the gas discharged from the equipment will rise, which is a normal phenomenon.
PACKAGE LIST
1x Oxygen generator host
1x Battery
1x AC adapter (with power cord)
1x Nasal oxygen tube
1x Portable package
ALARM INFORMATION
Alarm item |
Alarm reason |
Alarm Interval |
Screen display and prompts |
Processing method |
No breathing detected |
No breathing |
10s |
No breathing detected |
Please connect the nasal oxygen tube for use |
Low battery |
The battery power is between 0% and 20% when only using battery power |
10s |
Low battery |
Please charge in a timely manner |
Compressor shutdown |
Compressor not started or faulty |
10s |
Compressor |
Check whether the machine is overheated, and whether the air inlet and cooling holes are blocked.Restart the machine after 10 minutes of shutdown. Please contact the seller promptly |
System temperature too high |
The system temperature is higher than 60 degrees Celsius |
10s |
System temperature too high |
Check whether the air inlet and cooling holes of the machine are blocked. Restart the machine after 20 minutes of shutdown |
Fan stall |
The fan stops rotating |
10s |
Fan stall |
Alarm only, please contact the seller in a timely manner |