Instruction User Manual of RESPIREASY RE-S2L Oxygen concecntrator 1-9L DEDAKJ


 

 

 

 

 

 

PRODUCT USER MANUAL

HOMECARE OXYGEN CONCENTRATOR

MODEL NO: RE-S2L

 

 

 

 

BRAND INTRODUCTION

 

RE-S2L Oxygen Concentrator is designed and manufactured by Shenzhen Jiaxing Technology Innovation Co., Ltd.

RESPIREASY, developed by Shenzhen Jiaxing Technology Innovation Co., Ltd., has specialized in oxygen concentrators for over 15 years. We provide 1–60 LPM portable, home, and medical-grade devices with original accessories, serving 400,000+ users in 60+ countries.

Backed by German respiratory experts, we hold 100+ patents and focus on PSA efficiency, AI adaptive flow, and IoT monitoring. Our ISO 13485-certified Shenzhen factory (CE/ISO/NMPA) has a 2M annual capacity, with strict 7-step testing ensuring 93% purity and <0.5% failure rate in 2 years.

RESPIREASY delivers reliable, affordable, and smart oxygen therapy solutions for home, travel, and clinical care.

--RESPIREASY, easy respiratory care, smart oxygen therapy solutions

 

 

Intended Use of the Product

This is a home appliance that separates oxygen from air to produce oxygen -rich gas. This product is not a medical device and is not suitable for surgical, emergency, or critically ill patients.

Danger

· Smoking is strictly prohibited while using this product! Do not place matches , lit cigarettes , or any ignition sources in the room where the oxygen concentrator is located. 

Warning

·  Do not operate the product without reading or fully understanding this manual!

·  If you have any questions about the warnings or operating instructions ,

please contact respireasy.com for support. Do not operate the product on your own, as this may result in serious personal injury or property damage .

 

Attachment Warning

This oxygen concentrator is developed and produced by RESPIREASY company; please use it with the included accessories. Accessories produced by other manufacturers may affect the performance of this product; and are not recommended for use with the product

 

CONTENTS

 

Safety Precautions ------------------------------------------------- 1
Product Features --------------------------------------------------- 3
Product Structure -------------------------------------------------- 5
Installation and Use ------------------------------------------------ 6
Remote Control Usage --------------------------------------------- 7
Maintenance and Care --------------------------------------------- 8
Cleaning, Disinfection, and Sterilization --------------------------  9
Service Commitment ----------------------------------------------- 9
Common Troubleshooting Methods ------------------------------ 10
Handling, Transport, and Storage Conditions --------------------- 11
Precautions --------------------------------------------------------- 11
Packing List ---------------------------------------------------------  12


 

 

 Safety Precautions

Danger! Risk of electric shock. Do not disassemble the machine!

Please contact RESPIREASY for technical assistance .

 

Warning

Special care must be taken to reduce fire hazards when using oxygen . Any flammable and non-flammable materials can become highly flammable and burn rapidly in high oxygen concentrations . For safety, any flammable materials should be kept away from the molecular sieve oxygen concentrator , preferably not in the same room; and "No Smoking".

 

·  Oil , grease , or oily substances are highly susceptible to self-fuel when they come into contact with oxygen under pressure. These materials must be kept away from the oxygen concentrator , oxygen tubing , connectors , and other oxygen-producing components .

·  It  is recommended that you switch-on the oxygen generator for the best persormance for more than 30 minutes.Frequent machine switch-off fora short period time can reduce service lifespan .

·  Keep the power cord away from heat sources or hot surfaces!

·  Do not move the oxygen concentrator by pulling on the power cord!

·  Do not drop or plug foreign objects into any openings!

·  Do not block the air intake of the oxygen concentrator or place it on soft surfaces,such as beds or sofas that may obstruct airflow . Keep the air inlet away from lint,hai , or similar materials .

·  When adding water to the humidifier cup , follow the usage instructions and do not add excess water to avoid water spillage. Clean , pure water is recommend.

Place the oxygen concentrator in a well-ventilated indoor area , avoiding direct sunlight , and maintain at least 10 cm of space from walls , carpets ,    furniture , and similar objects .

·  It is recommended to use a anti-crush, registered standard nasal cannula .

·  The radio frequency energy used by this device only affects its internal operation; thus, its emissions are low and should not impact nearby electrical devices. However, severe interference from nearby RF-emitting devices or other electrical noise sources can disrupt its function.

·  If interference occurs , check the environment for the source and take measures to eliminate it:

(1)  Turn off nearby devices and then turn them back on;

(2)  Change the location of the interfering device;

Increase the distance between the interfering device and this equipment .

 

·  Do not use the oxygen concentrator while bathing. If continuous use is required , it must be placed in a room at least 2 meters away from the bathroom .

·  Avoid connecting with the oxygen concentrator when wet . Do not use or store the device in places where it may fall into water or other liquids . Do not touch a device that has fallen into water or other liquids! If it falls into water , unplug it immediately!

·  Always unplug the device when not in use!

·  If the power cord is damaged , it must be replaced by the manufacturer ’s service department or a qualified professional to avoid hazards .

 

 

Before Use

Unpacking Instructions: Unless the oxygen concentrator is to be used immediately, the machine must be kept in its original box and packaging materials before use.

· First, check the box or other packaging for any visible damage. If damaged, please notify the carrier or local dealer promptly.

· Inspect the exterior of the oxygen concentrator for any damage such as chips, dents, or scratches.

· Remove all bulk packaging materials from the box.

· Take out all components from the box and carefully check each one. Transportation and Storage Conditions

· Ambient temperature range: -20℃ to +55℃

· Relative humidity range: ≤93%, with no condensation.

· Atmospheric pressure range: 500 hPa - 1060 hPa

Note: The oxygen concentrator should be stored indoors, away from strong sunlight, corrosive gases, and in a well-ventilated area. During transportation, avoid severe vibration and do not transport it lying down or upside down.

Storage: Store the repackaged oxygen concentrator in a dry place. Do not place any objects on top of the oxygen concentrator.

 

 

② Product Features 

Ⅰ.Overview

The oxygen inhalation function: Inhaling oxygen can effectively improve the  body's oxygen supply, achieving the purpose of oxygen supplementation and  health maintenance. It is suitable for the elderly, pregnant women, those with weak constitutions, as well as students and office workers who experience varying degrees of physiological hypoxia. It can also be used to relieve fatigue  and quickly restore physical functions after intense physical or mental exertion.

This series of oxygen concentrators uses high-quality molecular sieve as the adsorbent, producing oxygen that meets health standards directly from the air  through the pressure swing adsorption (PSA) method. These oxygen concentrators are compact, lightweight, energy-efficient, low-noise, and easy  to operate.

 

Ⅱ.Scope of Application

This product is not a medical device and is intended solely for household and

personal oxygen production, not for life support or life extension.

 

Ⅲ. Key Technical Indexs

1. Brand: RESPIREASY;

2. Model: RE-S2L;

3. Oxygen Generation Method: Pressure Swing Adsorption (PSA)

4. Flow Range: 1-9Liter per minute;

5. Operating Mode: Continuous operation;

6. Oxygen Concentration Range: 93%-30%;

7. Overall Noise: 36±3dB(A);

8. Power Supply: AC110V/60Hz (please use the power cord provided with the

device);

9. Input Power: 120-170W;

10. Net Weight: 5.8KG;

11. Dimensions: 205×280×390mm;

12. Altitude: Oxygen concentration will not decrease from sea level to 1828 meters;

 

Safety System:

▶Power overload or loose connections will cause shutdown;

▶High temperature of the compressor will cause shutdown;

 

13. Time Function: ≥30 minutes;

14. Normal Operating Environment:

▶Ambient Temperature Range: 5℃~40℃;

▶Relative Humidity: ≤80%RH;

 

15. Atmospheric Pressure Range: 860hPa-1060hPa.

Note: When storage or transport temperature is below 5℃, the device should  be placed in a normal operating temperature environment for more than four hours before use.

16. Electrical Classification: Class II equipment;

17. Oxygen Outlet Temperature: ≤46℃;

18. Display Control: LED light display, Intelligent chip control. Infrared remote control

 

IV. Working Principle 

Air is compressed by an oil-free air compressor, and the compressed air flows to the inlet P of the rotary valve, which is controlled by the control panel.

 

1. When the rotary valve is energized, the compressed air flows from P to A, then from the A port of the rotary valve to the molecular sieve (A bed). The nitrogen in the compressed air is adsorbed by the molecular sieve, while oxygen enters the flow meter to control the output flow for oxygen use.  Another path enters the molecular sieve (B bed) through the desorption port for washing and desorption, and the desorbed gas is discharged to the atmosphere through the exhaust silencer of the rotary valve.

 

2. When the rotary valve loses power, the compressed air flows from P to B, then from the B port of the rotary valve to the molecular sieve (B bed). The nitrogen in the compressed air is adsorbed by the molecular sieve, while oxygen enters the flow meter to control the flow for oxygen use. Another path enters the molecular sieve (A bed) through the desorption port for washing and desorption, and the desorbed mixed gas is discharged to the atmosphere through the exhaust silencer of the rotary valve.

 

After this working cycle ends, the rotary valve is controlled again, entering the program for A. This process continuously adsorbs nitrogen and collects oxygen, enabling the two beds (A and B) to work alternately, thereby producing oxygen directly from the air and providing a continuous supply of oxygen for use.

 

③ Product Structure

 

Figure 1: (Display Screen Illustration)

 

 

 

 

 Front                                                                                          

RESPIREASY RE-S2L Oxygen concecntrator 1-9L

 

 Back             

RESPIREASY RE-S2L Oxygen concecntrator 1-9L

 

 

 

 Installation and Use

I. Preparation

1 . Remove the oxygen concentrator from the packaging box and take off all packaging materials .

 

2 . (As shown in Figure 5) Connect one end of the "oxygen delivery tube " to the   humidifier bottle interface and the other end to the oxygen outlet of the concentrator; you can also connect it directly to a nasal cannula for use .

 

Note: The humidifier bottle can function without water ; adding water is to humidify the oxygen and will not affect the oxygen production efficiency.

3. To add water to the humidifier bottle , loosen the bottle cap , add purified water , ensuring the water level does not exceed   the maximum water level line; then tighten the bottle cap and pull the elastic securing strap to place the humidifier bottle in the fixed slot of the oxygen concentrator .

 

4. Connect the power cord:

First , ensure that the power switch of the oxygen concentrator is off. Then connect one end of the power cord to the power socket of the oxygen concentrator and the other end to a grounded socket with power output . Do not use an extension cord.

 

 

 

 

 

 

 

Figure 5

RESPIREASY RE-S2L Oxygen concecntrator 1-9L

 

 

II. Oxygen Inhalation Operation

1. Turn the power switch to “I”, then press the “Oxygen Production” button on the control panel. The oxygen concentrator icon will light up, indicating it is in normal operation. While the concentrator is working, it will emit a “thud-- puff--” sound every few seconds, which is the normal sound of ventilation and exhaust. You can adjust the flow rate using the “Flow” button.

 

Note: The oxygen concentration will stabilize after approximately 30 minutes of initial operation.

 

2. As shown in Figure 6, connect one end of the nasal cannula to the oxygen outlet of the humidifier bottle and ensure the other end is properly worn by the patient to begin oxygen inhalation.

-- Adjust the oxygen inhalation time and flow rate according to medical advice.

-- After completing oxygen inhalation, turn off the oxygen concentrator and remove the oxygen delivery tube.

Figure 6

RESPIREASY RE-S2L Oxygen concecntrator 1-9L

 

 

III. Nebulization Port (Applicable for High-Configuration Model)

1、Before powering on, turn the nebulization port cap counterclockwise to  remove it.

2、Connect one end of the catheter to the nebulization port. Please use as  needed or under the guidance of a qualified physician, and wear the nebulizer as required.

3、To power on, press the “Oxygen Production” button. Ensure the flow rate is set to 1L; the oxygen machine will switch to nebulization mode.

4、After nebulization is complete, press the “Oxygen Production” button to power off, disconnect the nebulizer tubing, and reattach and tighten the nebulization port cap.

 

IV. Timer Setting

This machine has a timer function. If needed, press the “Timer” button on the display screen to adjust. The maximum timer duration is 4 hours (suggested based on actual circumstances).

 

V. Volume Adjustment: Press the volume button to adjust the sound level.

 

VI. Negative Ion Switch: Press the negative ion button to turn the negative ion function on or off.

 

VII. Flow Adjustment

Press the flow button to adjust the oxygen flow rate. Different settings correspond to different oxygen concentrations, which will be displayed on the panel (unit: L/min).

 

VIII. Power Off

·During use, if you need to stop oxygen inhalation, you can press and hold the “Oxygen Production” button on the control panel to stop or restart the oxygen supply.

·After use, first disconnect the oxygen delivery tube or connection from the humidifier outlet, turn off the power switch, and then unplug the power cord.

 

 Remote Control Usage

1.Power Button

2.Flow Rate Button

3.Timer Button

4.Negative Ion Button

    

RESPIREASY RE-S2L Oxygen concecntrator 1-9L

 

⑥ Maintenance and Care

Warning: Before Maintenance of the Oxygen Concentrator, Disconnect the Power Supply.

 

·The oxygen concentrator should be placed in a dry, dust-free, and well- ventilated area. Ensure that the air intake and exhaust ports are unobstructed.

· Oxygen delivery devices (such as nasal cannulas) should be used by designated individuals to ensure hygiene.

 

I.Cleaning the Entire Unit

The exterior of the unit should be wiped clean at least once a month. First, disconnect the power supply, then use a clean, soft, slightly damp cloth or sponge to wipe it down. Do not allow liquid to seep into the seams of the casing.

 

II. Cleaning or Replacing the Filter and Air Intake Sponge Filter

Cleaning and replacing the filters and air intake sponge filter is crucial for  protecting the compressor and molecular sieve, and for extending the lifespan of the oxygen concentrator. Please clean or replace them promptly.

* The oxygen filter should be replaced every 2-3 months

 

Warning:

·Do not operate the oxygen concentrator if the filter or air intake sponge filter  is not installed or is damp, as this can cause permanent damage to the machine.

1, Removing the Filter/Air Intake Sponge Filter:

The filter/air intake sponge filter is located at the back of the machine. Gently pull the filter outwards as indicated by the arrow in Figure 7(The actual cleaning time for the filter should be determined based on usage time and environmental conditions. If there is significant dust accumulation, it should be cleaned or replaced immediately, regardless of the duration of use.)

2, Cleaning &Replacing:

The expired oxygen filter may lead to DEDAKJ Oxygen home oxygen concentrator can not put out oxygen like it should be.

So the oxygen filter should be replaced the new one in time. 

The air filter should be replaced every 2-4 months, depending on the usage environment.

 

III. Cleaning the Humidifier Bottle

The water in the humidifier bottle should be changed daily, and the filter should be cleaned once a week. First, wash with a mild detergent, then rinse thoroughly with clean water to ensure hygienic oxygen use. When cleaning the humidifier bottle, be sure to also clean the accompanying humidifier cap.

 

Note: If the humidifier bottle has not been cleaned for a long time or if the water has not been changed, resulting in scale or odors, soak it in vinegar for at least half an hour.

Shake it to clean, then rinse thoroughly with clean water.

IV. Cleaning the Nasal Cannula

The nasal cannula is a single-use product. If reused, it should be cleaned after each use. Soak it in white vinegar for about 5 minutes, then rinse thoroughly with clean water, or wipe it with medical alcohol.

 

Note: The inside of the nasal cannula should be kept dry during use.

 

 

 Cleaning, Disinfection, and Sterilization

1. The oxygen delivery head can be wiped with alcohol after each use.

2. Generally, it should be cleaned once a week. Use a mild detergent for cleaning, and then rinse thoroughly with clean water.

 

Note: Ensure that the water in the liquid bottle and tubing is completely dried  before reuse.

 

⑧ Service Commitment

·From the date of purchase, this product is covered by a one-year free warranty for faults due to quality issues under normal use, and lifetime  maintenance is provided with this card.

·The free warranty does not cover the machine's casing and external  accessories.

·The interpretation rights of the terms related to this "Service Commitment" belong to Shenzhen Jiaxing Technology Innovation Co., Ltd.

The following situations are not covered by the warranty:

1.Consumables: disposable nasal cannulas, casters, sponge filters, filters, humidifier bottles.

2.Damage caused by liquids such as water or medications entering the machine  due to user actions.

3.Water damage from immersion or exposure to rain.

4.Faults caused by unauthorized disassembly, repair, or modification of the product.

5.Damage resulting from accidental drops during use or transport.

6.Faults caused by improper operation not following the instructions.

7.Damage caused by unforeseeable natural disasters (e.g., fire, earthquake, flood, etc.).

 

⑨Common Troubleshooting Methods

Fault Phenomenon

Fault Analysis

Handling Methods

Power switch is on , but the machine does not

work or the power

switch indicator is

not lit .

Poor contact between the power plug and socket .

Securely plug in the power cord .

The socket has no power output .

Move to a socket with power output .

Damage to the main board

Replace it

The machine

operates normally

with normal

sounds, flow rate

adjustments work ,

but the oxygen

output is very low

or nonexistent .

Check if the nasal can nula has any defects .

Replace the nasal cannula .

Ensure the humidifier bottle  and cap are properly sealed to prevent air leaks .

Tighten or replace the humidifier bottle cap .

Verify that the humidifier bottle  is securely attached to the machine .

Reinstall the humidifier bottle , ensuring  it  is flush with the machine 's installation surface .

 

 

After operating for a period, the Machine is overheated or shuts down .

.Air intake and exhaust ports are obstructed .

The oxygen  concentrator should    be placed in a well-ventilated indoor area, at leas 15 cm away from walls ,curtains, furniture, and similar objects .

The air intake filter is dirty .

Check  if the  air  intake  filter  sponge  on the back shell of  the oxygen concentrator is blocked or dirty, and clean it promptly .

The machine is overheating

The machine is equipped with a high - temperature protection device. If  it automatically shuts down due to excessive temperature,  turn off  the oxygen concentrator,  then check  if  the intake and exhaust ports are  blocked and if the filter sponge is dirty.  Wait for  the machine to cool down before restarting it .

 

Regular  and  slight "puffing "

exhaust sound .

Normal

The  sound of  the oxygen  concentrator extracting oxygen and expelling waste gas is a normal operating phenomenon .

Note:  If the fault cannot be  resolved using the above methods, please contact customer service. 

Email: office@respireasy.com

 

 

⑩ Handling, Transport, and Storage Conditions

I.carry

1.Unboxing

Note: Unless the oxygen concentrator is going to be used immediately, the cardboard box and storage packaging materials must be kept before using the machine.

·First, check the cardboard box or other packaging for any obvious damage. If there are any signs of damage, please notify the carrier or local dealer promptly.

·Remove all loose packaging materials from inside the box.

·Carefully take out all components from the packaging.

 

2.Inspection

·Check the exterior surface of the oxygen concentrator for any damage, such as chips, dents, or scratches.

·Inspect all components.

 

3.Storage

·Store the repackaged oxygen concentrator in a dry place.

·No objects should be placed on top of the oxygen concentrator.

 

II. Transportation and Storage Conditions

·Environmental temperature range: -20°C to +55°C

·Relative humidity range: ≤93% and without condensation

·Atmospheric pressure range: 500 hPa to 1060 hPa

Note: The oxygen concentrator should be stored indoors away from strong  sunlight, corrosive gases, and in a well-ventilated area. During transportation,  avoid severe vibrations and do not place it upside down or horizontally.

⑪ Precautions

· Oxygen is an oxidizing gas; please do not smoke and avoid open flames.

· Do not frequently power on the oxygen concentrator to prevent pressure-related failures that may affect its lifespan.

· If the power cord is damaged or the plug shows signs of potential hazards, do not use it.

· Ensure that the intake and exhaust ports of the oxygen concentrator are unobstructed, with at least 15 cm of clearance from any obstacles around the exhaust port.

· Please replace the water in the humidification bottle daily before using the oxygen concentrator. If not used for an extended period, empty the water from the humidification bottle.

· The oxygen concentrator should be used exclusively by one person to maintain hygiene.

· Do not use parts, accessories, or adapters that are not approved by the manufacturer. Using non-specified humidifiers and accessories may reduce the machine's performance.

· Dispose of waste and residues according to national regulations.

 

Packing List

Upon unboxing, please check against the following packing list. If any components are missing or damaged, promptly contact the sales representative  for replacement.

Packing List

Upon unboxing, please check against the following packing list. If any components are missing or damaged, promptly contact the RESPIREASY sales representative for replacement.

 

RE-P2L oxygen concentrator     x1 unit

Power cord                              x1 piece

Nasal oxygen tube                     x1 piece

Filter                                      x1 piece

Remote control                        x1 piece

User manual                             x1 piece

Certificate                               x1 piece

 

 