Steps of Use Oxygen Concentrator DEDAKJ DE-1A DE-1B(ddt-1a 1b)

DEDAKJ Household Oxygen Concentrator

 

Model: DE-1A/DE-1B/DE-1LW/DE-1BW

 

DEDAKJ DE-1A 1B oxygen concentratorsDEDAKJ DE-1A 1B oxygen concentrators

Main Technical Indexes

 

Product name: Household Oxygen Concentrator

Model: DE-1A/DE-1B/DE-1LW/DE-1BW

Oxygen production method: PSA method

Rated voltage: AC220V/50Hz AC110V/60Hz Input power: 120VA

Oxygen flow: 1-7L/min, adjustable

Oxygen concentration: 30-90%

Oxygen outlet pressure: 20-70kPa

Carbon dioxide: pass the test by the specified method Carbon monoxide: pass the test by the specified method pH value: pass the test by the specified method 

Shape identification: colorless odorless gas, with combustion supporting property

Negative oxygen ion: ≥8,000,000/cm³

Atomizing rate: ≥0.2mL/min (limited to DE-1LW/DE-1BW)

Running noise: ≤45dB (A weighted)

Electrical classification: Class II

Product dimensions: 340x180x310mm Net weight: 6.0Kg

Normal working conditions

Ambient temperature range: 5℃~40℃ Relative humidity range: ≤80%RH

Atmospheric pressure range: 86kPa-106kPa

※When the temperature is lower than 5℃ , the equipment shall be placed under normal working conditions for more than four hours before start-up  for use.

Transportation & storage conditions

Ambient temperature range: -20℃~60℃ Relative humidity range: ≤93%RH

 

 

Steps of Use Oxygen Concentrator DE-1A/DE-1B(ddt-1a 1b)

 

① Drainage of water tank

Drain the water tank as shown in the Figure on the left.

② Water injection

Open the rubber plug on the top of the water   tank and inject pure water (cold boiled water    or distilled water) to below the highest water    level, then install the rubber plug of water tank back to its original position (remember to install the rubber plug of water tank back to the original position, so as to avoid no oxygen output from the oxygen outlet during normal operation of the machine due to oxygen leakage from the water filling port). Then, align the humidification water tank with the outlet of the oxygen machine water tank, align it with the bottom clamping slot, and install the humidification water tank back to its original position.

※It should not be higher than the highest water level. Water injection is not necessary, and it is only for humidifying oxygen gas. You may not    inject water if you don't think the oxygen is dry.

 

③ Check "air filter"

Open the air inlet cover and check whether the "air filter" is installed in place and clean.

 

④ Connect the power line

As shown in the Figure on the left, insert the output end of    the power line into the power socket of the oxygen machine, and then insert the other end of the power line into the socket.

 

⑤ Connect the oxygen inhalation tube Insert the large-diameter end of the oxygen inhalation tube into the "oxygen outlet" of the host engine.

※ If there are too many water drops in the water collector during or after oxygen inhalation, please open the water collector and pour out water.

 

⑥Wear the oxygen inhaler

As shown in the Figure, when using the ear hook type oxygen inhaler, adjust the position of the oxygen outlet and align it with the nostrils; when using a snuffing oxygen inhaler, please align the two outlets with the nostrils and tighten the rubber ring.

 

⑦ Start the oxygen machine

Turn on the "power switch" and touch the "Start/Stop" key gently. Then, the machine will start to produce oxygen, and the user can inhale oxygen.

 

Please adjust the oxygen concentration to the corresponding gear according to the practical situation. Recommended oxygen concentration: 70%~90% for special people (please do not inhale high-concentration oxygen unless otherwise advised by the doctor) The oxygen concentra- tion of 50%~70% is suitable for the elderly and pregnant women, and 30%~50% is suitable for office workers and students, or please consult a physician.

 

 

Contact DEDAKJ Office